• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3 NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3 NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number BOST610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 09/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint (b)(4).A4: patient weight is not provided / unknown.E1: initial reporter¿s title and last name are not provided / unknown.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that when patient presented for second stage exposure surgery, we diagnosed with a localized advanced peri-implantitis defect characterized by gingival inflammation, exudate to probing with a 8-9mm probing depth and a small sequestrum of necrotic bone attached to the implant surface.The clinician surgically exposed the implant, removed the necrotic bone, tested the implant for osseointegration (passed torque test at 20 n/cm) and rescheduled the patient for re-evaluation.Site # 19 (universal).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 NON-PLATFORM SWITCHED TAPERED IMPLANT 6 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key19633096
Report Number0001038806-2024-01371
Device Sequence Number7400656
Product Code DZE
UDI-Device Identifier00844868029700
UDI-Public(01)00844868029700(17)270219(10)2022010795
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberBOST610
Device Lot Number2022010795
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-