| Catalog Number |
157001120 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Synovitis (2094); Swelling/ Edema (4577)
|
| Date of Event |
06/01/2010
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Pinnacle mom medical records received.After review of the medical records the patient was revised to address right hip pain with bursitis and alval.Operative notes reported copious yellow tinged clear fluid from the bursa, deficiency in the anterior abductors approx.20 %.There was inflame synovium and this was removed.There was a small amount of scuffing in the acetabulum and in the superior quadrant on the peripheral aspect of the bearing surface causing metal debris.Doi: (b)(6) 2008.Dor:(b)(6) 2010.Affected area: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.Product description:- summit por taper sz6 std off.Product code:- 157001120.Lot no:- cy1g11000.Quantity of manufactured:- (b)(4).Date of manufacturing:- jun-2008.Expiry date:- jun-2018.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot = > product description:- summit por taper sz6 std off.Product code:- 157001120.Lot no:- cy1g11000.Quantity of manufactured:- (b)(4).Date of manufacturing:- jun-2008.Expiry date:- jun-2018.Device history review: a manufacturing record evaluation was performed for the finished device, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|