• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INFINITY POLY SZ 2 8MM TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INFINITY POLY SZ 2 8MM TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33652208
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Implant Pain (4561); Swelling/ Edema (4577)
Date of Event 06/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient may need to undergo a revision surgery due to increased pain and swelling 6-8 months post-op.The patient's infection work up has been negative.
 
Additional Manufacturer Narrative
Device is not available for return as it remains implanted in the patient.When additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the patient may need to undergo a revision surgery due to increased pain and swelling 6-8 months post-op.The patient's infection work up has been negative.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.Upon further investigation of the ct scans, healthcare professionals opined that- ¿there is failure confirmed with the surgeon's assessment.Infection was considered as a potential underlying cause.The ct scan shows some radiolucence and small cysts.Loosening of the tibia is likely.There is no clear sign of migration.The talus looks unsuspicious.No comment can be given regarding the underlying cause.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.¿if device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY POLY SZ 2 8MM TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19635572
Report Number3010667733-2024-00406
Device Sequence Number8673379
Product Code HSN
UDI-Device Identifier00840420122636
UDI-Public00840420122636
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33652208
Device Lot Number1682090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2024
Supplement Date Received by Manufacturer10/22/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date11/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexUnknown
-
-