• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TAC123715A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Great Vessel Perforation (2152)
Date of Event 06/02/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
H6: code b20 ¿ the device remains implanted and, therefore, was not available for direct analysis by gore.H6: code c19-a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.H6: code a26- used to capture insufficient information available to determine if gore device contributed to stroke.It should be noted that, per the gore® tag® thoracic branch endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, aortic rupture and death.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2024, patient underwent an endovascular procedure to treat a thoracic aneurysm utilizing gore® tag® thoracic branch endoprosthesis and gore® tri-lobe balloon catheter.Upon attempting to balloon the proximal end of tac123715a device, the patient's aorta ruptured.Medical staff attempted cpr on the patient for 10 minutes, and ultimately patient could not be revived and expired.It was also reported there were no issues during the procedure, while implanting the devices.The balloon did not rupture and was not over inflated.In addition, no calcification was noted.Physician does not know what caused this issue.Physician also reports as patient was older, patient presented with a fragile aorta.No further information is available.
 
Additional Manufacturer Narrative
H6: code a26- used to capture insufficient information available to determine if gore device contributed to patient death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
lillian espinoza
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19635867
Report Number2017233-2024-05063
Device Sequence Number18295375
Product Code MIH
UDI-Device Identifier00733132654215
UDI-Public00733132654215
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P210032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/08/2026
Device Catalogue NumberTAC123715A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2024
Supplement Date Received by Manufacturer06/28/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age83 YR
Patient SexFemale
-
-