H6: code b20 ¿ the device remains implanted and, therefore, was not available for direct analysis by gore.H6: code c19-a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.H6: code a26- used to capture insufficient information available to determine if gore device contributed to stroke.It should be noted that, per the gore® tag® thoracic branch endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, aortic rupture and death.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2024, patient underwent an endovascular procedure to treat a thoracic aneurysm utilizing gore® tag® thoracic branch endoprosthesis and gore® tri-lobe balloon catheter.Upon attempting to balloon the proximal end of tac123715a device, the patient's aorta ruptured.Medical staff attempted cpr on the patient for 10 minutes, and ultimately patient could not be revived and expired.It was also reported there were no issues during the procedure, while implanting the devices.The balloon did not rupture and was not over inflated.In addition, no calcification was noted.Physician does not know what caused this issue.Physician also reports as patient was older, patient presented with a fragile aorta.No further information is available.
|