DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 173024
Medical Device Problem Codes Break (1069); Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: d1, d2, d4(model#, catalog#, expiration date, lot#, udi#), d8, g3, g4, h3, h4, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the needle was returned loaded in an endo stitch instrument and was removed for further inspection.The needle was broken at the suture attachment site.No suture was present with needle.It was reported that the device was difficult to load and the needle was broken.The reported issue were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the toggle levers are fully retracted prior to opening the jaws of the device.The reloading button should never be pressed when the instrument is in the body cavity, as this will release the needle.Inspect the application site to ensure hemostasis.Place additional sutures or use electrocautery if necessary to complete hemostasis.Endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.A thorough understanding of the operating principles, risks versus benefits, and the hazards involved in utilizing an endoscopic approach is necessary to avoid possible injury to the user and/or patient.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
D10 concomitant products: 73016 - 173016 endo stitch instrument, lot# j2d1482ey.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a thoracic procedure, the needle keeps breaking inside the handle.It was also reported that the device was difficult to load.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIDAC
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19636913
Report Number9612501-2024-01590
Device Sequence Number7093611
Product Code OCW
UDI-Device Identifier10884521129450
UDI-Public10884521129450
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number173024
Device Catalogue Number173024
Device Lot NumberJ2F0390Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2024
Supplement Date Received by Manufacturer08/26/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date09/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-