Catalog Number 150410104
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).E3 initial reporter occupation: lawyer this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2015 the patient underwent a left total knee arthroplasty involving attune components.The patella was resurfaced and smartset cement x 2 was used.On (b)(6) 2024, the patient underwent a left knee revision for instability.Intraoperatively, the tibial component was completely debonded at an unknown interface.The patient had severe osteolysis and a tibial plateau fracture.The fracture was repaired with a small trabecular metal cone and bone grafts.Competitor products were placed during the revision while the attune patella was left in situ.Doi: (b)(6) 2015 dor: (b)(6) 2024 (left knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM LT SZ 4 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19636933
Report Number1818910-2024-14300
Device Sequence Number7146179
Product Code JWH
UDI-Device Identifier10603295041610
UDI-Public(01)10603295041610(10)8079378(17)250228
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number150410104
Device Lot Number8079378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/18/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 3 CEM; ATTUNE MEDIAL DOME PAT 35MM; ATTUNE PS FB INSRT SZ 4 6MM; SMARTSET MV 40G - EO; SMARTSET MV 40G - EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-