Catalog Number 3122040
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2015 the patient underwent a left total knee arthroplasty involving attune components.The patella was resurfaced and smartset cement x 2 was used.On (b)(6) 2024, the patient underwent a left knee revision for instability.Intraoperatively, the tibial component was completely debonded at an unknown interface.The patient had severe osteolysis and a tibial plateau fracture.The fracture was repaired with a small trabecular metal cone and bone grafts.Competitor products were placed during the revision while the attune patella was left in situ.Doi: (b)(6) 2015 dor: (b)(6) 2024 (left knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET MV 40G - EO
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19637667
Report Number1818910-2024-14303
Device Sequence Number17915599
Product Code LOD
UDI-Device Identifier10603295168379
UDI-Public(01)10603295168379(10)7996598(17)161030
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K081155
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2016
Device Catalogue Number3122040
Device Lot Number7996598
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/18/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/08/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 3 CEM; ATTUNE MEDIAL DOME PAT 35MM; ATTUNE PS FB INSRT SZ 4 6MM; ATTUNE PS FEM LT SZ 4 CEM; SMARTSET MV 40G - EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-