H6: code b20 ¿ the device remains implanted and, therefore, was not available for direct analysis by gore.H6: code c19-a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.It should be noted that, per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2023, patient underwent an endovascular procedure to treat a ruptured aneurysm utilizing gore® excluder® aaa endoprosthesis.On an unknown date, imaging follow-up revealed there was a type ii endoleak.On (b)(6) 2024 (gore associate was not present), patient underwent reintervention surgery, and the ima was coiled.Another follow-up scan (unknown date), revealed there was a persistent type ii endoleak from the lumbar arteries.On (b)(6) 2024, an angiogram was performed.Coil embolization of the two lumbar arteries was performed by additional coiling of the aaa sac.No type 1b or issue was observed on the plc181200 contralateral leg component, but access was gained into the aaa sac, and right common iliac artery limb was extended to maximize the seal zone as a precautionary measure.Patient tolerated the procedure.No images are available.
|