BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SERUM; EVACUATED BLOOD COLLECTION TUBE IVD, CLOT ACTIVATOR
|
Back to Search Results |
|
| Catalog Number |
367820 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® serum tube, the tube did not clot leading to the sample not being able to be used.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples, but 3 photos were provided for investigation.The photos were reviewed and the indicated failure mode for missing additive was observed.Additionally, 30 retention samples from bd inventory were evaluated by visual examination and the issue of missing additive was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode missing additive.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|