| Model Number |
SUTURELASSO, 45 DEGREE CURVE RIGHT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It as reported that during a labrum refix surgery the tip of the device broke off.The surgeon was able to retrieve the broken pieces.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-4068-45r serial/batch number (b)(6) was received for investigation.Functional testing was not performed as the device tip was broken off.Visual inspection revealed that the device's tip had broken off at the weld, likely due to excessive force applied during use.The observed condition is likely due to prying/levering the device.
|
| |
|