ZIMMER BIOMET, INC. KINECTIV MODULAR FEMORAL STEM PRESS-FIT CEMENTLESS SIZE 16.25; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Hematoma (1884); Failure of Implant (1924)
Date of Event 06/10/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial hip procedure.Subsequently, the patient was revised approximately 3 months later due to periprosthetic fracture following a fall.During the revision 5 cerclage wires were used to reduce and stabilize the fracture.The stem and head were replaced without complications.It was reported that no further information was available.
 
Additional Manufacturer Narrative
(b)(4).D10: cat# 00784801300; lot# 66285602; modular neck p 12/14 neck taper use with +0 heads only.Cat# 00620206022 lot# 63835146 shell porous with cluster holes 60 mm.Cat# 00630506040; lot# 66229505; liner standard 3.5 mm offset 40 mm i.D.For use with 60 mm o.D.Shell.Cat# 00625006530; lot# j7593906; bone screw self-tapping 6.5 mm dia.30 mm length.Cat# 00625006530; lot# j7618988; bone screw self-tapping 6.5 mm dia.30 mm length.Cat# 110005096; lot# 0002551062; juggerknot 2.9 w/ tapered ndls.Cat# 110005096; lot# 0002548224; juggerknot 2.9 w/ tapered ndls.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: g3; h2; h3; h6.Proposed component codes: mechanical (g04) stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the patient underwent an initial right tha.The patient experienced a fall and was revised as it resulted in a periprosthetic fracture.During the revision, the stem was noted to be loose.A definitive root cause cannot be determined for the loosening and fracture.It was reported the patient fell, however the reason for the fall is unknown.No problem was found with the device in relation to the hematoma after review by a hcp.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINECTIV MODULAR FEMORAL STEM PRESS-FIT CEMENTLESS SIZE 16.25
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19643583
Report Number0001822565-2024-02181
Device Sequence Number7734556
Product Code LPH
UDI-Device Identifier00889024132689
UDI-Public(01)00889024132689(17)330421(10)65850316
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K182678
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00771301600
Device Lot Number65850316
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2024
Supplement Date Received by Manufacturer11/26/2024
Initial Report FDA Received Date07/01/2024
Supplement Report FDA Received Date12/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight81 KG
-
-