| Model Number |
620-2431411-01 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device would begin to charge, but then internally dumped the energy.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to hv capacitor.The hv capacitor was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|