Model Number SYNERGY RESECTION SHAVER CONSOLE
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 06/12/2024, it was reported by a sales representative via (b)(4) that an ar-8305 shaver console is causing an issue.This is one of five devices that are linked together.When plugged in, one or more of these devices cause a loss of power for all instruments using that same outlet (they are plugged into a medical grade outlet on the wall in the or, not just a power strip).The contact is unable to figure out which causes the issue; therefore, they'd like to replace them all.This occurred during use in a case with no patient effect.
 
Additional Manufacturer Narrative
(no problem found) the evaluation did not identify any issues relevant to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY RESECTION SHAVER CONSOLE
Common Device Name
POWERED SURG ORTHOPEDIC INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19644055
Report Number1220246-2024-06522
Device Sequence Number12700662
Product Code HWE
UDI-Device Identifier00888867123854
UDI-Public00888867123854
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K932699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSYNERGY RESECTION SHAVER CONSOLE
Device Catalogue NumberAR-8305
Device Lot Number15049114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2024
Supplement Date Received by Manufacturer07/08/2024
Initial Report FDA Received Date07/01/2024
Supplement Report FDA Received Date07/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/08/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-