| Model Number |
EEA33 |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Adverse Event Without Identified Device or Use Problem (2993); Firing Problem (4011)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
|
| Date of Event |
06/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the treatment of deeply infiltrating endometriosis, despite proper installation and the appearance of the green indicator, when the stapler was fired, there was a very high resistance, which prevented full compression of the handle.There was a deformation of the staples and a blood loss of 200ml.A section of the intestine was cut off and another circular stapler from the same batch was used with no problem.Due to the device issue, the patient was fitted with a temporary protective stoma for three months.Surgical time was extended for an hour.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was an issue with staple formation, difficult to fire, and staple line bleeding.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|