• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. NEXGEN 15MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. NEXGEN 15MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent a left knee replacement approximately twelve months post-implantation due to loosening.No additional patient consequences were reported.
 
Additional Manufacturer Narrative
(b)(4).D10 medical devices: nexgen prc agmt block post sz f 10mm catalog#: 00599003602 lot#: 64068372.Nexgen prc agmt block post sz f 10mm catalog#: 00599003602 lot#: 63149149.Nexgen distal femoral augment block precoat size f 5 mm augment with screw catalog#: 00599003610 lot#: 64746753.Nexgen distal femoral augment block precoat size f 5 mm augment with screw catalog#: 00599003610 lot#: 65323179.Rotating hinge knee left size f cemented option femoral catalog#: 00588001601 lot#: 64862679.Size 4 precoat cemented tibial component catalog#: 00588000400 lot#: 65901693.Trabecular metal tibial cone, medium 31x31mm catalog#: 00545001331 lot#: 64882569.Tibial half block augment tapered precoat size 4 15 mm catalog#: 00598800428 lot#: 63383185.12mm diameter 100mm length straight stem extension catalog#: 00598801012 lot#: 64822355.All-poly patella cemented 38 mm diameter catalog#: 42540200038 lot#: 65885881.Articular surface with segmental hinge post size f 17 mm height catalog#: 00585006017 lot#: 64990221.17mm height size f articular surface with hinge post extension catalog#: 00588006017 lot#: 64495430.G2 foreign source: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, h10, and h11.Proposed component code: mechanical (g04) stem.Medical records were reviewed but the reported issue was unable to be confirmed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: femoral component not completely seen but no definite loosening clearly visualized.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN 15MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19645326
Report Number0002648920-2024-00212
Device Sequence Number15391446
Product Code JWH
UDI-Device Identifier00889024220997
UDI-Public(01)00889024220997(17)300630(10)64796921
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00598801015
Device Lot Number64796921
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2024
Supplement Date Received by Manufacturer11/06/2024
Initial Report FDA Received Date07/01/2024
Supplement Report FDA Received Date11/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight74 KG
-
-