• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL S.R.O. GENTLECATH OTHER; CATHETER, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL S.R.O. GENTLECATH OTHER; CATHETER, UROLOGICAL Back to Search Results
Model Number 501014
Medical Device Problem Code Device Misassembled During Manufacturing /Shipping (2912)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
End user reports "the indicator line on the funnel not aligned with the coudé tip.This makes the use of the coudé catheter more difficult.In order to use the catheters, he is marking the catheter on the funnel end with a line acting as the indictor line for the coudé tip.".
 
Additional Manufacturer Narrative
(b)(6).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 3005778470.
 
Additional Manufacturer Narrative
Review of the dhr showed that all relevant tests required during the manufacturing process, packaging process and final product release had been fulfilled and met the requirements.No nonconformity has been registered during the manufacturing process of the mentioned lot.No another complaint was received to this lot.The batch record review indicates no discrepancies.Should additional information become available, a follow-up report will be submitted.Fda registration number.Reporting site: 1049092.Manufacturing site: 3005778470.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLECATH OTHER
Common Device Name
CATHETER, UROLOGICAL
Manufacturer (Section D)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO  07101
Manufacturer (Section G)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO   07101
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key19646057
Report Number3005778470-2024-00889
Device Sequence Number7751418
Product Code KOD
UDI-Device Identifier00768455145539
UDI-Public00768455145539
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K172924
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number501014
Device Lot Number3C03866
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2024
Supplement Date Received by Manufacturer07/19/2024
Initial Report FDA Received Date07/01/2024
Supplement Report FDA Received Date07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-