MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ROD FOR REAMING GUIDE HOLDER; ORTHOPEDIC INSTRUMENT HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ROD FOR REAMING GUIDE HOLDER; ORTHOPEDIC INSTRUMENT HANDLE Back to Search Results
Catalog Number 230774002
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the front part of the rod for reaming guide holder instrument is bent and would not thread into the holder for reaming guide.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: this instrument is composed of 1 part.The front part of the rod for reaming guide holder instrument is bent and would not thread into the holder for reaming guide( part no.Provided for reference only).The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed the tip of rod for reaming guide holder slightly bent to one side.It is reasonable to confirm a difficulty to assemble the device, as condition may had happen as a consequence of the observed damage.Potential cause can be attributed to unintended use error by use of excessive force and overloading the material.Properly handling and attention to the approved use of the device diminishes the risk of failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the rod for reaming guide holder would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device 5259518 number, and no non-conformances nor manufacturing irregularities were identified.Device history review: a manufacturing record evaluation was performed for the finished device 5259518 number, and no non-conformances nor manufacturing irregularities were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: this instrument is composed of 1 part.The front part of the rod for reaming guide holder instrument is bent and would not thread into the holder for reaming guide (part no.Provided for reference only).The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed the tip slightly deformed of the rod for reaming guide holder.Additionally, device had signs of usage.Potential cause can be attributed to unintended use error by use of excessive force and overloading the material.Properly handling and¿attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was unable to be performed as the mating component was not returned for evaluation.However, it is reasonable to confirm a difficulty to assemble the device, as condition may had happen as a consequence of the observed damage.The overall complaint was confirmed as the observed condition of the rod for reaming guide holder would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device 5259518 number, and no non-conformances nor manufacturing irregularities were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: this instrument is composed of 1 part.The front part of the rod for reaming guide holder instrument is bent and would not thread into the holder for reaming guide( part no.Provided for reference only).The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed the tip of rod for reaming guide holder slightly bent to one side.It is reasonable to confirm a difficulty to assemble the device, as condition may had happen as a consequence of the observed damage.Potential cause can be attributed to unintended use error by use of excessive force and overloading the material.Properly handling and attention to the approved use of the device diminishes the risk of failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the rod for reaming guide holder would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device 5259518 number, and no non-conformances nor manufacturing irregularities were identified.Device history review : a manufacturing record evaluation was performed for the finished device 5259518 number, and no non-conformances nor manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD FOR REAMING GUIDE HOLDER
Common Device Name
ORTHOPEDIC INSTRUMENT HANDLE
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19646125
Report Number1818910-2024-14406
Device Sequence Number6624791
Product Code PHX
UDI-Device Identifier10603295116134
UDI-Public(01)10603295116134(10)5259518
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K170748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number230774002
Device Lot Number5259518
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2024
Supplement Date Received by Manufacturer09/02/2024
09/23/2024
10/24/2024
11/12/2024
Initial Report FDA Received Date07/01/2024
Supplement Report FDA Received Date09/03/2024
09/25/2024
10/25/2024
11/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-