EXACTECH, INC. GXL NUETRAL LINER, G0 28MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
130-28-50 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Scar Tissue (2060); Osteolysis (2377); Bursitis (4832)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant devices: 4766349 142-28-03 - cocr fem head 28mm +3.5 offset 12/14.4793043 164-13-10 - novation element ro s/o col sz 10.4923699 180-65-20 - alteon 6.5mm screw, 20mm.4949119 180-01-46 - cup, cluster-hole, 46mm group 0.The product associated with the reported event is within the scope of recall z-1731-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was the cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 79 months after a left total hip replacement procedure, the patient underwent a revision procedure to address prosthesis wear, wear and delamination of the acetabular liner, and a small osteolitic lesion in the superior acetabulum.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: h6 (health effect - clinical code, type of investigation).The most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|