| Model Number |
V173 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode and premature battery depletion was suspected.The physician opted to explant and replace this device.No additional adverse patient effects were.This device has not returned for analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode and premature battery depletion was suspected.The physician opted to explant and replace this device.No additional adverse patient effects were.This device has not returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption (during telemetry or other higher power device operations) and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
| |
|