• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's rear case to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted stryker to report that their device's battery pin were broken.In this state, defibrillation therapy may be delayed or prevented from being delivered to the patient if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19651991
Report Number0003015876-2024-02054
Device Sequence Number6595925
Product Code MKJ
UDI-Device Identifier00883873871683
UDI-Public00883873871683
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2024
Initial Report FDA Received Date07/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-