| Model Number |
CC60WF |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that during an intraocular lens (iol) implant procedure, lens had a scratch noticed after implantation.The technician who was in charge of assembling the lens indicates that she did not notice if the lens had a scratch when she looked at it or when she opened and mounted it.Surgeon did not remove the lens and replace it, left the same lens on the patient and finished the surgery.Additional information was received, there was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6 and h.11.The product was not returned.The photo provided shows the lens implanted in the eye.A mark is observed along the edge of the optic of the lens.We are unable to determine the extent of damage to the lens from the photo.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.Based on our observation of the attached photos, it appears the lens has a mark on the optic.It is difficult to make a final determination without evaluation of the physical sample.The root cause cannot be determined based on available information.Associated products were not provided.It is unknown if qualified products were used.Company foldable iols (intra ocular lens) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu (instructions for use) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Company lens ifu: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to request for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|