| Model Number |
U-0572 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported via medwatch mw5155061 that, "two pieces of taka suction came apart while in use, both retrieved, no harm to patient." no report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Steris requested the device subject of the reported event be returned for evaluation.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The taka suction tube was returned to the supplier for evaluation.Evaluation results determined that excessive force was applied to the device.When the user applied excessive force at the connection of the tube and the handle, it caused the device to break as stated in the reported event.Steris performed in-service training on the proper use and operation of the taka suction tube, specifically the importance of not bending or applying excessive force while in use.The instructions for use states, "avoid mechanical shock or overstressing the devices." "prior to use, inspect devices to ensure proper function and condition.Do not use devices if they do not satisfactorily perform their intended function or if they have physical damage." the device history record was reviewed and confirmed the lot was manufactured according to specification; no abnormalities were noted.No additional issues have been reported.
|
| |
|