| Catalog Number |
367862 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.The date received by manufacturer has been used for this field.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd vacutainer® k2 edta (k2e) plus blood collection tubes have had multiple instances of tube push off.There has been no report of patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary material #: 367862; lot/batch #: 3318858.Bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and 10 tubes were selected for functional testing.No issues were observed relating to tube push off as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode tube push off.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd vacutainer® k2 edta (k2e) plus blood collection tubes have had multiple instances of tube push off.There has been no report of patient impact.
|
| |
|