| Catalog Number |
L20313 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Fatigue (1849); Hypersensitivity/Allergic reaction (1907); Necrosis (1971); Scar Tissue (2060); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
|
| Date of Event |
09/12/2012
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # = >(b)(4).E3 initial reporter occupation: lawyer this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Litigation and medical record received.After review of the medical records it was reported that the patient was revised to address bilateral adverse reaction to metal debris, elevated metal ions, cysts in trochanteric region and pseudotumor.Clinical visit reported muscle fatigue, cyst and some lucency on his femoral component.Operative note reported large fluid collection around hips encountered.There were large cystic lesion along with some necrotic tissue.There was a hypertrophic reaction in the synovium.These were debrided and hip reduction were performed.However, there were no metallosis.Metal liner and head were revised to ceramic on ceramic.Doi: may 25, 2007; dor: sep 12, 2012; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|