DEPUY ORTHOPAEDICS INC US AML 15.0 STD 6.3 5/8 STD 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
155405000 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Ambulation or Postural Difficulties (2544); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
|
| Date of Event |
10/07/2012
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received ad 18 jun 2024.After review of medical records patient was revised was due to adverse reaction to metal debris right hip.Prior to revision patient mri showed failed pseudotumor but she had a large fluid collection at the trochanter with avulsion of the anterior glutei.Pain has been stable but intermittent in severity and complaints audible grinding.There is a slight limp.Operative notes indicated upon removing the head there was a tremendous amount of metallic, black debris at the trunnion.Doi: (b)(6) 2006.Dor: (b)(6) 2012.Right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Product description:- aml 15.0 std 6.3 5/8 std 12/14, product code:- 155405000, lot no:- z21ca1000, quantity of manufactured:- (b)(4), date of manufacturing:- 15-aug-2005, expiry date:- 31-jul-2015.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : product description:- aml 15.0 std 6.3 5/8 std 12/14, product code:- 155405000, lot no:- z21ca1000, quantity of manufactured:- (b)(4), date of manufacturing:- 15-aug-2005, expiry date:- 31-jul-2015.Device history review : a manufacturing record evaluation was performed for the finished device, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|
|
|