W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
ATAA43120160A |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
06/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2024, the physician was treating a thoracoabdominal aneurysm with a gore® excluder® thoracoabdominal branch endoprosthesis (tambe).After implantation, while reviewing the final angiography, the physician noticed a dissection extending 3-4cm proximally from a primary entry tear just proximal to the tambe.The physician treated by covering the entry tear with a ctag device, but commented that intramural hematoma extended to the lsa.Reportedly, the physician was trying to stent an accessory renal, and was lasering through the graft, and also used a cutting balloon to make fenestrations, pulling the graft down while the proximal anchors were engaged.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation findings code c20: the device remains implanted and is not available for analysis.H.6.Investigation findings code c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.B.5.Description of event: added information about why physician was making fenestrations in the graft.
|
| |
|
Event or Problem Description
|
|
Reportedly, the physician made fenestrations in the device because there was an additional accessory renal that he wanted to reach.
|
| |
|
Search Alerts/Recalls
|
|
|