Model Number 8309EL
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: the device was received in and inspected by the manufacturer.Visual inspection did not note any damages to the alarm.A sensor was returned for evaluation but not the one involved in the incident as that was disposed of by the facility.Functional testing performed on the alarm was unable to confirm the reported issue of the unit not sounding.The alarm was found to have met specifications and passed all functional testing.The sensor that was returned showed issues when used with the alarm plugged into the ac power.It is believed that the issue is with the sensor and not the alarm itself.Based on the ability to turn the issue on and off through manipulation of the sensor neck, the issue is due to damage within the neck.Likely cause being mishandling during shipment to the customer or use by the customer.The instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.The ifu states that to reduce the risk of serious injury or death, test the alarm and sensor for proper operation prior to putting in service with a patient, and each time before leaving the patient unattended.If the alarm and/or sensor do not function properly, remove the alarm and sensor from service and replace them with a properly functioning alarm and/or sensor.Do not use the alarm or sensor if it does not activate each time weight is removed from the sensor or the sensor belt is unfastened.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Event or Problem Description
Customer reporting a complaint on product # 8645.Customer states that a patient was being monitored by sensor & 8645 alarm.Patient got up the alarm did not sound.Patient suffered a fall and laceration to their head.The 8645 s/n (b)(6) is available for return but sensor was disposed of by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARM SENSOR, CHAIR SENSOR PAD SQUARE 30-DAY
Common Device Name
MONITOR, BED PATIENT
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key19656025
Report Number2182318-2024-00095
Device Sequence Number12399472
Product Code KMI
UDI-Device Identifier10190676004369
UDI-Public10190676004369
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8309EL
Device Catalogue Number8309EL
Device Lot Number4073T122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2024
Initial Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-