XTANT MEDICAL HOLDINGS, INC. STREAMLINE MIS SPINAL FIXATION SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 05-FINALDRIVER2
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient underwent an initial two (2) level transforaminal lumbar interbody fusion procedure with posterior fixation using an mis approach.During the final tightening step in the procedure, it was noted that the set screw may have been tightened beyond the torque limit, causing the distal tip of the driver to fracture off.No pieces were retained by the patient.A spare final driver was used to finish the remainder of the surgery.No patient complications were reported.No additional information was available.
 
Additional Manufacturer Narrative
The reported event was unable to be confirmed due to limited information received from the complainant.No device was returned for evaluation; further no photographs were provided.Operative notes and/or medical records were not provided for review of usage/technique.It was noted that the surgeon tightened the device beyond the torque limit.It is unknown if a torque limiting handle or hand tightening was used when the malfunction occurred.A review of manufacturing records for the device was performed and no deviations were identified.The device was released to the field in november 2019, resulting in an approximate field life of over four (4) years.No further investigation was able to be completed at this time.A definitive root cause was unable to be determined with the information provided; however, may have been related to excessive force applied to the instrument.The ifu for the streamline mis system warns "instruments are subject to damage during use as well as long-term potentially damaging effects such as wear.Damage may result in significant risks to safety and/or inability to function as intended.If instruments are damaged or broken during use, metal fragments can be viewed by radiographic assessment.It is the surgeon's responsibility to carefully consider the risks and benefits of retrieving the fragments".Additionally, the ifu instructs the user "prior to and during use, including reprocessing, inspect instruments for: damage such as but not limited to, wear, discoloration, corrosion, cracking, fracture, or unrecognizable markings.Proper function including but not limited to, sharpness, movement of hinges and couplings, joint stability, and legible markings.Instruments that show signs of damage or an inability to function should not be used and should be returned to the manufacturer".If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STREAMLINE MIS SPINAL FIXATION SYSTEM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer (Section G)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT 59714
Manufacturer Contact
rebecca lennemann
664 cruiser lane
belgrade, MT 59714
4069245878
MDR Report Key19656481
Report Number3005031160-2024-00021
Device Sequence Number12418064
Product Code LXH
UDI-Device Identifier00846468088453
UDI-Public00846468088453
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number05-FINALDRIVER2
Device Lot Number351536
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2024
Initial Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-