The following sections were updated/corrected updated: b4, b5, g3, g6, h1, h2, h3, h6, h10, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: office notes 1st follow up: anterior knee pain, 'has flat back syndrome and walks with his knees in a flexed position'; office notes 2nd follow up: chronic knee pain after replacement of right knee joint.Trouble with ambulation and rom.Complaint is confirmed with the provided medical records.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|