| Model Number |
FHC-111 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event: was entered as 1jun2024 as the event date was not provided.Results pending completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported receiving false positive hcg results while using fisher sure-vue hcg urine dipsticks on an unknown number of patients with two lots.Each patient presented for a pregnancy check prior to performing an unspecified non-emergent diagnostic procedure.All of the patients provided a fresh urine sample that was tested immediately.The results were read at 4 minutes and positive hcg results were obtained.As a result, the diagnostic procedures were delayed 24 hours as confirmatory blood testing was performed.No adverse events were reported.No further information was provided.Per the package insert, "allow the test strip, urine specimen and/or controls to equilibrate to room temperature (15-30°c) prior to testing." see mdr 2027969-2024-00088 for the event on lot number 0000695546.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 - date of event: was entered as 1jun2024 as the event date was not provided.D4 - model #: was updated to fhc-111 and was previously submitted as fhc-101.Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
|
| |
|
Event or Problem Description
|
|
The customer reported receiving false positive hcg results while using fisher sure-vue hcg urine dipsticks on an unknown number of patients with two lots.Each patient presented for a pregnancy check prior to performing an unspecified non-emergent diagnostic procedure.All of the patients provided a fresh urine sample that was tested immediately.The results were read at 4 minutes and positive hcg results were obtained.As a result, the diagnostic procedures were delayed 24 hours as confirmatory blood testing was performed.No adverse events were reported.No further information was provided.Per the package insert, "allow the test strip, urine specimen and/or controls to equilibrate to room temperature (15-30°c) prior to testing." see mdr 2027969-2024-00088 for the event on lot number 0000695546.
|
| |
|