• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIAC SCIENCE CORPORATION POWERHEART AED G3 PLUS AUTOMATIC AED; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIAC SCIENCE CORPORATION POWERHEART AED G3 PLUS AUTOMATIC AED; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 9390A-1001SP
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly self-tests, on/off current, patient and circuit impedance testing, and shock testing without duplicating the report.An internal inspection found no discrepancies.The main board has been replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERHEART AED G3 PLUS AUTOMATIC AED
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
CARDIAC SCIENCE CORPORATION
500 burdick parkway
deerfield WI 53531 9692
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key19664817
Report Number2112020-2024-00513
Device Sequence Number17777157
Product Code MKJ
UDI-Device Identifier00812394021093
UDI-Public(01)00812394021093
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number9390A-1001SP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2024
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/18/2024
Initial Report FDA Received Date07/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-