• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER GOLD F3 25MM STER; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER GOLD F3 25MM STER; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A0411225G0303
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that a protaper gold f3 25mm ster file broke.The broken part remains in the canal.
 
Additional Manufacturer Narrative
Summary: returned protaper gold finishing file f3 25mm is actually broken in the active part (fatigue).No material defect was found during analysis of the rupture pattern.No unused file is available for evaluation.The batch number is unknown, dhr cannot be reviewed.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTAPER GOLD F3 25MM STER
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key19665015
Report Number8031010-2024-00076
Device Sequence Number17800179
Product Code EKS
UDI-Device IdentifierD716PTGRF3251
UDI-PublicD716PTGRF3251
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 07/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA0411225G0303
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2024
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/21/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/21/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/03/2024
Supplement Report FDA Received Date07/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-