• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; INTERNA NARROW PLATFORM IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; INTERNA NARROW PLATFORM IMPLANT Back to Search Results
Model Number IIPECA3310
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes High Blood Pressure/ Hypertension (1908); Unspecified Infection (1930); Inflammation (1932); Bruxism (4427)
Date of Event 10/30/2020
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician indicates that he placed the implant on (b)(6) 2020 and, during the second surgery (four months later) it was observed that the implant had not osseointegrated.Patient with a bone quality iv.In the event the patient experienced: infection and pain.
 
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
INTERNA NARROW PLATFORM IMPLANT
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street, 14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street, 14
minano menor, alava 01510
SP   01510
Manufacturer Contact
jose rivero
leonardo da vinci street, 14
minano menor, alava 01510
SP   01510
MDR Report Key19666449
Report Number3004417597-2024-00012
Device Sequence Number6904812
Product Code DZE
UDI-Device Identifier08435389813156
UDI-Public08435389813156
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K211952
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date02/01/2025
Device Model NumberIIPECA3310
Device Lot NumberBO00411.Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2020
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/23/2020
Initial Report FDA Received Date07/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-