| Model Number |
IRPUCA3710 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994)
|
| Date of Event |
10/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The clinician indicates that he placed the implant on (b)(6) 2020 and, during the second surgery (two months later) it was observed that the implant had not osseointegrated.Patient with a bone quality iii.In the event the patient experienced: infection and pain.
|
| |
|
Additional Manufacturer Narrative
|
|
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.However, there is a small gap between the implant and the cover screw, as well as signs of infection around this area.
|
| |
|