DEPUY SYNTHES PRODUCTS LLC 1.12MM X 6.35MM FRTR,S; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
S-6R-G1 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the trigger of the burr device drill bit was broken off inside and could not be removed.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: d1, d2 d4, g4, h4: the brand name, common device name, procode, lot number, catalog, udi, expiration date, 510(k) and device manufacture date have been updated accordingly.Investigation summary the actual device was returned for evaluation.During repair, an evaluation was performed; and it was determined that the device had the following failures: cannot remove cutter and corrosion/rusting/pitting.Therefore, the reported condition that a drill bit device was broken off inside the attachment device and could not be removed was confirmed.The assignable root cause was determined to be due to improper handling which is user error.
|
| |
|
Search Alerts/Recalls
|
|
|