| Catalog Number |
3542625 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Breast Discomfort/Pain (4504)
|
| Date of Event |
06/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: breast pain and deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female who underwent a breast augmentation primary with mentor siltex round moderate profile, 200cc saline prosthesis experienced bilateral breast pain and bilateral deflation postoperative.The issue was discovered during in office visitation.As a result, patient scheduled for bilateral removal on (b)(6) 2024.This report relates to the left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received on july 24, 2024, indicated that the patient underwent bilateral implant replacements as follow: left sn: (b)(6), right sn: (b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturing record evaluation (mre) was reviewed, and no anomalies were found related to this complaint.In addition, the mre review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on aug 07, 2024: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the sal siltex rnd diap pkg 200cc breast implant had parallel lines of shell wear on both views.In addition, the breast implant was found to be ruptured, received in two (2) parts, the valve was received completely separated.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the continuous and sustained stresses on the device.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Manufacturer¿s reference number: (b)(4).
|
| |
|