| Model Number |
LF-TP |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the intubation fiberscope's insertion section, soft tube marking disappeared.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over (19) years since the subject device was manufactured.Based on the results of the investigation, olympus presume that the defect was caused by stress of repeated use, external factors, or handling.The instruction manual ¿chapter 3 preparation and inspection, 3.2 preparation and inspection of the endoscope¿ describes the methods for inspection on the suggested event.Olympus will continue to monitor field performance for this device.
|
| |
|