• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 0.3ML 31G 8MM WHOLEUNIT 10BG 500; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 0.3ML 31G 8MM WHOLEUNIT 10BG 500; SYRINGE, PISTON Back to Search Results
Model Number 328438
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported liquid at the tip of needles.He said he noticed it when he depressed the plunger rod.Lot #: 3254919 catalog #: 328438 date of event: 06-10-24 samples: available - sending mail kit.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 2nd complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was not able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.3ML 31G 8MM WHOLEUNIT 10BG 500
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key19669343
Report Number3024508819-2024-00162
Device Sequence Number14244525
Product Code FMF
UDI-Device Identifier00382908438032
UDI-Public00382908438032
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K212499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328438
Device Lot Number3254919
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/03/2024
Supplement Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-