ABBOTT VASCULAR INC. NC TREK NEO¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
| Model Number |
1400250-20 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.The investigation determined that the reported balloon rupture appears to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is likely that the balloon interacted with the mildly tortuous and mildly calcified anatomy such that the balloon became damaged and subsequently ruptured during inflation.Based on the reported information, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported the 2.50 x 20mm nc trek neo rx was to be used in the mid left anterior descending (mlad) lesion.However, post dilatation, the balloon ruptured during the first inflation to 12 atmospheres (atm).Therefore, the balloon was removed and a non-abbott balloon was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|