• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. AVL ANGLED REAMER SLEEVE ASSEMBLY, 15; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. AVL ANGLED REAMER SLEEVE ASSEMBLY, 15; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number AVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 6/6/2024, a sales representative reported via sems-06587988 that an ar-9676 angled reamer, drive shaft, and ar-9297-15 univers vaultlock augmented glenoid angled reamer sleeves (part of set ar-9579s mgs augmented glenoid instrument set) were stuck together.This was discovered after the case during the dismantling of the setup, with no reported patient harm.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-9297-15 serial/batch number (b)(6) was received for investigation.Functional testing was not performed due to the returned ar-9676 angled reamer shaft being stuck inside the ar-9297-15.Visual inspection noted signs of wear on the device.The most likely cause is attributed to misuse due to interference with the mating instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19671350
Report Number1220246-2024-06581
Device Sequence Number3136920
Product Code LXH
UDI-Device Identifier00888867350038
UDI-Public00888867350038
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAVL ANGLED REAMER SLEEVE ASSEMBLY, 15
Device Catalogue NumberAR-9297-15
Device Lot Number052043
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2024
Supplement Date Received by Manufacturer09/11/2024
Initial Report FDA Received Date07/03/2024
Supplement Report FDA Received Date09/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-