| Catalog Number |
912068 |
| Medical Device Problem Codes |
Positioning Failure (1158); Fracture (1260); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): 912068 (0002323013).E1: (b)(6) medical center.G2: foreign - the event occurred in japan.The customer has indicated that the product is in the process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during an elbow collateral ligament suturing procedure, the tip of two (2) separate inserters bent and fractured during implantation of the anchor into the patient's burr hole.The surgeon noted that the drill positioning may have shifted, resulting in misalignment and excessive stress load.The fractured tip from this product was not retained by the patient.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The reported event is confirmed as the product was returned.Visual examination of the returned product revealed that it was returned without sutures, anchors, or needles.One side of the tip was fractured off and was not returned.The opposite side remains attached but is bent outward.The tip is also curved upward.A review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|