The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the objective lens fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the objective lens.In addition, our technician confirmed that the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the lcb (light carrying bundle) fluid damage, the objective lens cloudy (not clear view), the control body corroded, the operation channel (primary) leak, and the remote control buttons cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|