| Catalog Number |
8065752451 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that actuating failure of probe was confirmed during vitrectomy surgery.The condition of aspirating was unknown.The surgery was completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria the returned sample was visually inspected.Visual inspection confirmed that the aspiration line of the unused probe manifold was bent at the tubing insertion to the probe engine.The defect prevented fluid from flowing from the probe to the cassette.The most likely root cause for the customer¿s event is a manufacturing defect.The vitrectomy probe has a rear shell meant to improve the ergonomics of the handpiece, however, if the rear shell is improperly installed, it will result in the customers reported event.After an investigation of this complaint, it has determined that no further actions will be pursued at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|