• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAMILTON MEDICAL AG HAMILTON-T1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAMILTON MEDICAL AG HAMILTON-T1 Back to Search Results
Model Number 1610090
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Hamilton medical ag case number is: (b)(4).
 
Event or Problem Description
The following was reported to hamilton medical ag: during service, unable to calibrate pressure, does not hold pressure and expels it through the expiratory valve.No patient involvement.
 
Additional Manufacturer Narrative
Hamilton medical ag case number is: (b)(4).Hamilton medical ag complaint number: (b)(4).Follow-up 1 - corrected information: **udi related data quality updates only** field d4 was updated with full udi information.Updated fields: b4, d1, d3, d4, g6, h2, h4, h11.
 
Event or Problem Description
The following was reported to hamilton medical ag: during service, unable to calibrate pressure, does not hold pressure and expels it through the expiratory valve.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON-T1
Common Device Name
HAMILTON-T1
Manufacturer (Section D)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, 7402
SZ  7402
Manufacturer (Section G)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, 7402
SZ   7402
Manufacturer Contact
christof mueller
via crusch 8
bonaduz, 7402
SZ   7402
MDR Report Key19678015
Report Number3001421318-2024-01602
Device Sequence Number20541971
Product Code CBK
UDI-Device Identifier07630002813549
UDI-Public(01)07630002813549(21)25158(11)221117(240)1610090
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K181216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1610090
Device Catalogue Number1610090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2024
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date07/05/2024
Supplement Report FDA Received Date10/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-