• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 56MM O.D. SIZE KK POROUS UNCEMENTED WITH CLUSTER HOLES SHELL USE WITH KK LINERS; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 56MM O.D. SIZE KK POROUS UNCEMENTED WITH CLUSTER HOLES SHELL USE WITH KK LINERS; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that intraoperatively, a titanium cup (it cup) of 56 was first used, and then an attempt was made to implant the corresponding polyethylene liner, and it was found that the polyethylene liner and the cup of 56 could not be locked, so the cup was removed and an attempt was made to implant the polyethylene liner outside of the body, and it was found that it could not be locked, and then the surgeon went back to using the larger file, and implanted a cup of 58, and punched the 58 corresponding polyethylene liner, and it was not able to be locked.The doctor had no choice but to remove it, increase the size of the file, and finally used a 60 cup and the corresponding liner to lock it in place.It was reported to cause a sixty (60) minute delay.No other harm to the patient.
 
Additional Manufacturer Narrative
(b)(4).D10: 00875201232 32mm i.D.Size kk elevated rim liner use with 56mm o.D.Size kk shell 65603763 00875305801 58mm o.D.Size ll porous uncemented with cluster holes shell use with ll liners 65524646 00875201332 32mm i.D.Size ll elevated rim liner use with 58mm o.D.Size ll shell 65714273 g2: china customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Shell item 00875305601 lot 65540700 shows scratches to the rim face, inner spherical taper wall, and bio debris in fiber metal from attempted use.Visual product identifiers for fiber metal, (3) cluster holes, and (12x) anti rotation scallops were confirmed to print.Dimensional product identifier for overall diameter is in specification.Review of the device history records identified no deviations or anomalies during manufacturing.Review of complaint history found no additional related issues for these items and the reported part and lot combinations.Medical records were not provided.A definitive root cause cannot be determined.Complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
56MM O.D. SIZE KK POROUS UNCEMENTED WITH CLUSTER HOLES SHELL USE WITH KK LINERS
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19679995
Report Number0001822565-2024-02248
Device Sequence Number15166526
Product Code LPH
UDI-Device Identifier00889024150218
UDI-Public(01)00889024150218(17)320819(10)65540700
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151448
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00875305601
Device Lot Number65540700
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2024
Supplement Date Received by Manufacturer12/02/2024
Initial Report FDA Received Date07/05/2024
Supplement Report FDA Received Date12/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE
Patient Age62 YR
Patient SexMale
-
-