• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH RIA 2 BONE HARVESTING KIT 520MM STERILE; REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RIA 2 BONE HARVESTING KIT 520MM STERILE; REAMER Back to Search Results
Catalog Number 03.404.000S
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11: additional narrative: h3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date the patient underwent hardware removal after right humerus non-union.During the procedure, bone harvest clogged into the assembly.The rai 2 shaft was stuck inside the assembly; the plastic was cut away form the shaft.This report is for an rai 2 bone harvesting kit.This is report 3 of 5 for (b)(4).(b)(4) captures the intra-operative rai 2 issues.Related complaint (b)(4) captures the required revision procedure due to non-union.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIA 2 BONE HARVESTING KIT 520MM STERILE
Common Device Name
REAMER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19680332
Report Number8030965-2024-08281
Device Sequence Number6965367
Product Code HTO
UDI-Device Identifier10886982274236
UDI-Public10886982274236
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K993335
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.404.000S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2024
Initial Report FDA Received Date07/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
15.5MM REAMER HEAD FOR RIA 2 STERILE; 15.5MM REAMER HEAD FOR RIA 2 STERILE; 16.5MM REAMER HEAD FOR RIA 2 STERILE; RIA 2 REAMING KIT 520MM STERILE
Patient Age68 YR
Patient SexMale
-
-