| Catalog Number |
ER8075 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced due to a reservoir tubing tear.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced due to a reservoir tubing tear.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and inlet tubes of the pump.The surfaces of the detachment site of the inlet tube of the pump appear to be rough and irregular, indicating stress was exerted.Abrasion was noted on both exhaust tubes of the cylinders.No functional abnormalities were noted with either cylinder 1 or cylinder 2.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.While no functional abnormalities were noted during testing with the returned components, microscopic examination of the surfaces of the exhaust tube detachment end between the pump and reservoir revealed rough and irregular surfaces, indicating stress was exerted.Based on the overlapping of the pump tubes, in combination with device usage over time, this could contribute to sufficient stress to separate the inlet tubing of the pump.A separation of this type would then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Search Alerts/Recalls
|