| Model Number |
NA-U401SX-4022 |
| Medical Device Problem Code |
Problem with Sterilization (1596)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the aspiration needle experienced hole/irregularity in the sealed sterile package.The issue occurred during a diagnostic endobronchial ultrasound procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than a year (4 months) since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's reportable malfunction was not confirmed since the device was not sent back in the original packaging.Based on the results of the investigation, it is not possible to establish the root cause.Olympus will continue to monitor field performance for this device.
|
| |
|