• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PXC201000
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 06/10/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The fsa reported the following to gore: on (b)(6) 2009, the patient underwent an endovascular procedure utilizing the gore® excluder® aaa endoprostheses (2 contralateral legs, 1 aortic extender, and 1 trunk-ipsilateral leg).On (b)(6) 2024, during a follow up ct scan, the physician contacted the fsa and reported that the patient had bilateral type 1b endoleaks.They plan to implant one ibe most likely on the left limb and then possibly one on the right as well in a few weeks.
 
Additional Manufacturer Narrative
According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and / or require intervention include endoleaks.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H.6: code b15 - images were provided, and an imaging evaluation was performed.An analysis of relevant data will be performed in view of supporting the identification of possible causes of the event.The imaging evaluation determined the following: the post-implantation cta was dated (b)(6) 2024.There was contrast outside the implanted devices in the distal abdominal aorta aneurysm sac.Imaging showed the distal end of the contralateral leg on the patient¿s right-side ends in the abdominal aorta aneurysm.Thereby, there is lack of all apposition in the right common iliac (rci).The contralateral leg that extends from the ipsilateral leg into the left common iliac (lci) has also lost circumferential device apposition in the lci.Findings confirmed distal type 1 endoleaks, bilaterally.
 
Additional Manufacturer Narrative
Added g3/g4 and h1/h2 correction contact office - manufacturing site: from: (b)(6).To: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
monique nwanze
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19688498
Report Number3013164176-2024-02126
Device Sequence Number20865815
Product Code MIH
UDI-Device Identifier00733132610037
UDI-Public00733132610037
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date03/31/2012
Device Catalogue NumberPXC201000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2024
Supplement Date Received by Manufacturer08/07/2024
09/03/2024
Initial Report FDA Received Date07/08/2024
Supplement Report FDA Received Date08/22/2024
09/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/16/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexMale
-
-