On (b)(6) 2024, patient underwent an endovascular procedure to treat a pau in the descending thoracic aorta, utilizing gore® tag® conformable thoracic stent graft with active control system (ctagac).During the procedure, the graft had moved distally (15-20cm), due to physician inadvertently not releasing stage 3 (lockwire release) of the ctagac.There was a temporary coverage of the celiac, sma and both renals.The procedure was successfully completed by pulling the graft down further into the abdominal aorta and additional stents (palmaz stents) were placed inside.Patency of all of those vessels were restored and confirmed via angiogram.Patient tolerated the procedure.No images or further patient information is available.
|
|
H6: code c19-a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.The device remains implanted and, therefore, was not available for direct analysis by gore.Per the gore® tag® conformable thoracic stent graft with active control instructions for use (ifu) states the following: "the device remains attached to the catheter via the lockwire immediately following deployment to full diameter.Do not move the catheter or handle as unintended device movement or vessel damage may occur." it should be noted that, per the gore® tag® conformable thoracic stent graft with active control instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, branch vessel occlusion/obstruction and component migration.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|